IG3000PH/ORP
IG3000PH analyzer sensor is a digital composite electrode composed of precious metal indicator electrode and reference electrode. PH measurement is based on the measuring electrode and reference electrode of the working cell in solution measured potential difference, using the PH value of the solution to be measured and the working cell potential between the linear relationship, to achieve PH online monitoring.

Technical Parameter

Sensor:
Range: pH: 0~14pH; Temperature: 0 ~ 95 ℃
Measurement accuracy: ±0.01PH; Plus or minus 0.1 ℃
Zero potential PH: 7±0.25 PH (15mV)
Resolution: 0.01PH
Maximum pressure: ≤ 0.6mpa
Slope: 95% or higher
Resistance: 250 m or less Ω
Alkali error: 0.2pH(1mol/L Na+ pH14) (25℃)
Response time ≤10 seconds (95% end point value) (after stirring)
Calibration method: one-point or two-point buffer calibration (automatic identification)
Sensor material: Ryton, Ag/Agcl reference gel
Sensor voltage: 9~36VDC
Sensor communication: MODBUS RS-485
Medium temperature: 0~95℃ (no icing)
Cable length: 10 m cable is standard and can be extended
Sensor size: diameter 25.5mm* length 198mm
Weight: 0.5 KG

Controller:
Display: high contrast LCD large screen, with strong backlight can operate under direct sunlight, menu navigation
Controller material: ABS17/16-L3(body), PC17/16-L3 (protective cover)
Number of sensor channels: 4 channels
Output: 2-way 4-20mA (expandable)
Relay: 2 channels relay, program to set response parameters and response values
Digital communication: Optional MODBUS RS485 communication function, real-time transmission of measured values
Data storage: password type data storage and curve display function, monitoring parameter log type change dynamic
Storage temperature: -20 ℃ to 60℃
Operating temperature: -15~60℃
Protection level: IP66
Power supply: AC power supply: 90-240VAC (50/60Hz) Dc power supply: 24VDC
Dimensions: 188mm*160mm*106mm (length * width * height)
Weight: 1.45 KG


Market / Application
● Municipal wastewater ● Industrial wastewater ● Environmental monitoring ● Drinking water and surface water monitoring
● Industrial processes ● Aquaculture ● Biological processes


Characteristic of instrument
◆ Built-in temperature sensor, automatic temperature compensation
◆ Automatic identification standard liquid, 2 point calibration
◆ Digital sensor, RS-485 output, support standard ModbusRTU protocol, no
Affected by transmission distance, better stability
◆ The sensor is made of Ryton material for longer service life and reduced maintenance and downtime
◆ Large capacity salt bridge for longer probe life and less maintenance
◆ PH with flat bubble sensing element, optimized design to increase pollution resistance


E-mail: Germany@igenu.com  Tel: 49 (0) 9 35 89 00 24  Fax: 49 (0) 9 35 89 00 15